TRUE (1), used FPA index is valid
FPA_INVALID_NO (-2 or 0xFFFFFFFE) : FPA not opened with Multi API-DLL Functions | F_OpenInstancesAndFPAs or index out of range
IMPORTANT
...
BYTE : current FPA index as set using F_Set_FPA_index
5. F_Check_FPA_index
Anchor | ||||
---|---|---|---|---|
|
General Description
Get current FPA index and check if index is valid. A valid index corresponds to an individual FPA that has been opened with F_OpenInstancesAndFPAs. If FPA index was set to 0 (all), then this function will only return 0, even if no FPAs are open. Does not indicate whether FPA index is enabled or disabled.
...
current FPA index as entered using F_Set_FPA_index if selected FPA has been opened with Multi API-DLL Functions | F_OpenInstancesAndFPAs
FPA_INVALID_NO (-2 or 0xFFFFFFFE) : FPA not opened with Multi API-DLL Functions | F_OpenInstancesAndFPAs or index out of range
6. F_Enable_FPA_index
...
INT_X type return value from last function call on selected FPA
FPA_INVALID_NO (-2 or 0xFFFFFFFE) : FPA not opened with Multi API-DLL Functions | F_OpenInstancesAndFPAs or index out of range
9. F_Multi_DLLTypeVer
...
Code Block // return Type + Version // Type 0x6000 - Multi-FPA FlashPro430 // 0x7000 - Multi-FPA GangPro430 // 0x8000 - Multi-FPA FlashPro-CC // 0x9000 - Multi-FPA GangPro-CC // 0xC000 - Multi-FPA FlashPro2000 // 0xD000 - Multi-FPA GangPro2000 // 0x10000 - Multi-FPA FlashPro-ARM // 0x11000 - Multi-FPA GangPro-ARM // 0x17000 - Multi-FPA FlashPro-M // 0x18000 - Multi-FPA GangPro-M // Version = 0x0FFF & version
FPA_INVALID_NO (-2 or 0xFFFFFFFE) : FPA not opened with Multi API-DLL Functions | F_OpenInstancesAndFPAs or index out of range
10. F_Get_FPA_SN
...
SN of selected FPA
FPA_INVALID_NO (-2 or 0xFFFFFFFE) : FPA not opened with Multi API-DLL Functions | F_OpenInstancesAndFPAs or index out of range
11. F_Get_FPA_Label
...
updated FPA_LABEL structure pointed to by label input parameter
FPA_INVALID_NO (-2 or 0xFFFFFFFE) : FPA not opened with Multi API-DLL Functions | F_OpenInstancesAndFPAs or index out of range
Example of printing FPA_LABEL:
...
value between 0 and 100
FPA_INVALID_NO (-2 or 0xFFFFFFFE) : FPA not opened with Multi API-DLL Functions | F_OpenInstancesAndFPAs or index out of range
Example
...
PROG_OPCODE_VERIFY_ACCESS (1) : FPA is attempting to perform action described in F_Verify_Access to MCU.
PROG_OPCODE_FLASH_ERASE (2) : FPA is attempting to perform action described in F_Memory_Erase.
PROG_OPCODE_FLASH_BLANK_CHECK (3) : FPA is attempting to perform action described in F_Memory_Blank_Check.
PROG_OPCODE_FLASH_SELECTED_BLANK_CHECK (4) : FPA is attempting to perform action described in F_Memory_Blank_Check when a subset of the memory space is selected in configuration settings.
PROG_OPCODE_FLASH_WRITE (5) : FPA is attempting to perform action described in F_Memory_Write.
PROG_OPCODE_FLASH_VERIFY (6) : FPA is attempting to perform action described in F_Memory_Verify.
PROG_OPCODE_FLASH_READ (7) : FPA is attempting to perform action described in F_Memory_Read.
PROG_OPCODE_LOCK_MCU (8) : FPA is attempting to perform action described in F_Lock_MCU.
PROG_OPCODE_UNLOCK_MCU (9) : FPA is attempting to perform action described in F_Clear_Locked_Device.
PROG_OPCODE_START_APP (10) : FPA is attempting to run application programmed onto target device (if enabled to be ran after F_AutoProgram in configuration settings).
FPA_INVALID_NO (-2 or 0xFFFFFFFE) : FPA not opened with Multi API-DLL Functions | F_OpenInstancesAndFPAs or index out of range
Example
Insert excerpt | ||||||
---|---|---|---|---|---|---|
|
15. F_
...
CancelAction
General Description
Function that will attempt to cancel any running Encapsulated Functions , like the GUI Cancel button. This function is thread-safe. One thread can call F_AutoProgram (for example) and another thread can call F_CancelAction to stop the first. A reinitialization is recommended using F_OpenInstancesAndFPAs after calling this function.
...